Background Patient-ventilator dyssynchrony is common and may influence patients' outcomes. Detection of such dyssynchronies relies on careful observation of patients and airway flow and pressure measurements. Given the shortage of specialists, critical care nurses could be trained to identify dyssynchronies. Objective To evaluate the accuracy of specifically trained critical care nurses in detecting ineffective inspiratory efforts during expiration. Methods We compared 2 nurses' evaluations of measurements from 1007 breaths in 8 patients with the evaluations of experienced critical care physicians. Sensitivity, specificity, positive predictive value, negative predictive value, and the Cohen κ for interobserver agreement were calculated. Results For the first nurse, sensitivity was 92.5%, specificity was 98.3%, positive predictive value was 95.4%, negative predictive value was 97.1%, and κ was 0.92 (95% CI, 0.89-0.94). For the second nurse, sensitivity was 98.5%, specificity was 84.7%, positive predictive value was 70.7%, negative predictive value was 99.3%, and κ was 0.74 (95% CI, 0.70-0.78). Conclusion Specifically trained nurses can reliably detect ineffective inspiratory efforts during expiration.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.