Objectives. We aimed to study the feasibility and reliability of focused ultrasound (US) examinations to quantify pericardial (PE)- and pleural effusion (PLE) by a pocket-size imaging device (PSID) performed by nurses in patients early after cardiac surgery. Design. After a 3-month training period, with cardiologists as supervisors, two nurses examined 59 patients (20 women) with US using a PSID at a median of 5 days after cardiac surgery. The amount of PE and PLE was classified in four categories by US (both) and chest x-ray (PLE only). Echocardiography, including US of the pleural cavities, by experienced cardiologists was used as reference. Results. Focused US by the nurses was more sensitive than x-ray to detect PLE. The correlations of the quantification of PE and PLE by the nurses and reference was r (95% confidence interval) 0.76 (0.46–0.89) and 0.81 (0.73–0.89), both p < 0.001. PE and PLE were drained in one and six (eight cavities) patients, all classified as large amount by the nurses. Conclusions. Cardiac nurses were able to obtain reliable measurements and quantification of both PE and PLE bedside by focused US and outperform the commonly used chest x-ray regarding PLE after cardiac surgery.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.