-Prediction of bodily motion due to respiration was investigated preparatory to implementation of active compensation for respiration in a robot-assisted system for percutaneous kidney surgery. Data for preliminary testing were recorded from the chest wall of a subject using an optical displacement sensor. The weighted-frequency Fourier linear combiner algorithm, an adaptive modeling algorithm, was used to model and predict respiratory movement. Preliminary results are presented, in which the algorithm is shown to track a 0.86 mm rms respiration signal with 0.11 mm rms error. The general robotic system and compensation scheme are also described.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.