Purpose:
To evaluate long-term (4.5 to 6 y) changes in the position of Ahmed glaucoma valve (AGV) tubes.
Materials and Methods:
Adult patients aged 18 years and above, 1 to 2 months post-AGV implantation were enrolled. Tube position was evaluated using anterior segment optical coherence tomography at baseline and final follow-up. The intracameral tube length (ICL), tube-cornea (TC) distance, and cornea-tube angle were measured. Wilcoxon-signed rank test was used to assess whether the changes in parameters from baseline were significant.
Results:
In total, 11 eyes of 9 patients were included in the analysis. Mean duration of follow-up was 5.2±0.8 years (4.4 to 6.1 y). The ICL increased from 1.58±0.40 to 1.74±0.62 mm (P=0.18). The TC decreased from 0.36±0.12 to 0.29±0.05 mm (P=0.04). The cornea-tube angle decreased from 27.76±5.57 to 24.82±5.69 degrees (P=0.08). Overall, 73% of the tubes moved toward the cornea (mean 0.11 mm, over 4.4 to 6.1 y). In total, 73% showed an increase in ICL; 45% showed an increase of >0.3 mm [mean 0.50 mm (33%) increase]. Two of 11 tubes (18%) showed no significant change in the parameters (ICL, 0.03 and 0.01 mm increase; TC, 0 and 0.01 mm increase); these tubes were noted to lie on the iris.
Conclusions:
Tubes tend to move toward the cornea over time. To avoid corneal damage and involvement of the visual axis in the future, tubes should be reasonably short and inserted tangentially, preferably in the posterior one third of the anterior chamber.
The aim of this study was to measure the intraocular pressure (IOP) by Perkins hand held tonometer, non contact tonometer(NCT) and rebound tonometer with the Gold standard GAT and to analyse their correlation with varied CCT. Materials and Methods: Cross-sectional, hospital-based study that included 200 patients from the glaucoma department. After complete eye evaluation, the IOP measurement was performed sequentially in the same order-NCT, RT, Perkin's and GAT. CCT was then measured for all the patients. The intraocular pressure within the eyes was compared by paired "t" test and between the right and left eyes were compared by the independent "t" test. Results: Mean age of the patients was 54.5±13.5 years. The mean IOP of Right and Left eyes measured with GAT were 15.9±6.2 and 16.3±7.0 mm/Hg, that with Perkins were 15.7±6.2 and 16.3±7.0 mm/Hg respectively. Both eyes' mean IOP were 15.4±6.7 and 15.9±7.4 mm/Hg by Icare. The mean IOP measured by NCT were 16.4±6.7 and 16.9±8.0 mm/Hg. The mean CCT of right eye was 528.9± 29.6 and left eye was 530.6±29.6 microns. All the four instruments namely GAT, Perkin's, Icare ic100 and NCT were influenced by CCT. Conclusions: IOP remains the only alterable factor in the management of glaucoma patients. Numerous methods have been devised to measure IOP accurately. The IOP measured with Perkin's, NCT and Icare are comparable but GAT remains the gold standard.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.