This study reports the effect of exposure to liquid carbon dioxide on the mechanical properties of selected medical polymers. The tensile strengths and moduli of fourteen polymers are reported. Materials were exposed to liquid CO(2), or CO(2) + trace amounts of aqueous H(2)O(2), at 6.5 MPa and ambient temperature. Carbon dioxide uptake, swelling, and distortion were observed for the more amorphous polymers while polymers with higher crystallinity showed little effect from CO(2) exposure. Changes in tensile strength were not statistically significant for most plastics, and most indicated good tolerance to liquid CO(2). These results are relevant to evaluating the potential of liquid CO(2)-based sterilization technology.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.