The aim of this study was to develop an experimental set-up and a methodology to uniformly contaminate several filter samples with high concentrations of cultivable bacteria and fungi. An experimental set-up allows contaminating simultaneously up to four filters for range of velocities representative of heating, ventilating and air-conditioning systems. The test aerosol was composed of a microbial consortium of one bacterium (Staphylococcus epidermidis) and one fungus (Penicillium oxalicum) and aerosol generation was performed in wet conditions. Firstly, the experimental set-up was validated in regards to homogeneity of the air flows. The bioaerosol was also characterized in terms of number and particle size distribution using two particle counters: optical particle counter Grimm 1.109 (optical diameters) and TSI APS 3321 (aerodynamic diameters). Moreover, stabilities of the number of particles generated were measured. Finally, concentrations of cultivable microorganisms were measured with BioSamplers (SKC) downstream of the four filters.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.