Midogas® (CIG Australia) nitrous oxide/oxygen blenders are commonly used in delivery wards in Australia. In this case report we describe an incident where a small hole in the diaphragm of the oxygen failure alarm in a Midogas nitrous oxide/oxygen blender led to retrograde flow of oxygen contaminating the nitrous oxide pipeline supply to the operating theatres and the delivery ward. This caused a reduced level of nitrous oxide to the patient in theatre, but there was no adverse outcome. However, if the oxygen pipeline pressure had been lower than the nitrous oxide pipeline pressure, the contamination would have been reversed, potentially resulting in a hypoxic mixture being delivered to many patients. Suggestions are made as to a method of prevention of this problem.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.