These results demonstrate that the NeoChord procedure is safe, effective and reproducible. Clinical and echocardiographic efficacy is maintained up to 1 year with significant differences among the anatomical groups. Specific anatomical selection criteria are necessary to achieve stable results.
Atrial fibrillation (AF) is a common complication of cardiac surgery, with an increasing incidence. Post-operative AF results in many complications and increased healthcare resources. Despite substantial interest in the prediction and prevention of post-operative AF, as well as guidelines for the management of this common arrhythmia, there is still some uncertainty about appropriate risk stratification and management. The aim of this review article is to provide an overview of clinical predictive features for the development of AF following cardiac surgery and suitable preventive measures, using both antiarrhythmic and non-antiarrhythmic strategies.
Off-pump transapical implantation of artificial chordae to correct MR is technically safe and feasible; however, it yields further potential for improvement of efficacy and durability. (Safety and Performance Study of the NeoChord Device [TACT]; NCT01777815).
Background:
The tissue selectivity of pulsed field ablation (PFA) provides safety advantages over radiofrequency ablation in treating atrial fibrillation. One-shot PFA catheters have been shown capable of performing pulmonary vein isolation, but not flexible lesion sets such as linear lesions. A novel lattice-tip ablation catheter with a compressible 9-mm nitinol tip is able to deliver either focal radiofrequency ablation or PFA lesions, each in 2 to 5 s.
Methods:
In a 3-center, single-arm, first-in-human trial, the 7.5F lattice catheter was used with a custom mapping system to treat paroxysmal or persistent atrial fibrillation. Toggling between energy sources, point-by-point pulmonary vein encirclement was performed using biphasic PFA posteriorly and either temperature-controlled irrigated radiofrequency ablation or PFA anteriorly (RF/PF or PF/PF, respectively). Linear lesions were created using either PFA or radiofrequency ablation.
Results:
The 76-patient cohort included 55 paroxysmal and 21 persistent atrial fibrillation patients undergoing either RF/PF (40 patients) or PF/PF (36 patients) ablation. The pulmonary vein isolation therapy duration time (transpiring from first to last lesion) was 22.6±8.3 min/patient, with a mean of 50.1 RF/PF lesions/patient. Linear lesions included 14 mitral (4 RF/2 RF+PF/8 PF), 34 left atrium roof (12 RF/22 PF), and 44 cavotricuspid isthmus (36 RF/8 PF) lines, with therapy duration times of 5.1±3.5, 1.8±2.3, and 2.4±2.1 min/patient, respectively. All lesion sets were acutely successful, using 4.7±3.5 minutes of fluoroscopy. There were no device-related complications, including no strokes. Postprocedure esophagogastroduodenoscopy revealed minor mucosal thermal injury in 2 of 36 RF/PF and 0 of 24 PF/PF patients. Postprocedure brain magnetic resonance imaging revealed diffusion-weighted imaging+/fluid-attenuated inversion recovery- and diffusion-weighted imaging+/fluid-attenuated inversion recovery+ asymptomatic lesions in 5 and 3 of 51 patients, respectively.
Conclusions:
A novel lattice-tip catheter could safely and rapidly ablate atrial fibrillation using either a combined RF/PF approach (capitalizing on the safety of PFA and the years of experience with radiofrequency energy) or an entirely PF approach.
Registration:
URL:
https://www.clinicaltrials.gov
; Unique identifiers: NCT04141007 and NCT04194307.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.