A monoclonal antibody-based flow-through immunoassay (FTA) was developed using a nitrocellulose membrane placed on the top of adsorbent pads enclosed in a plastic cassette with a test zone at the center. The FTA could be completed within 10 min. Clear purple dots against a white background indicated the presence of Aphanomyces (A.) invadans. The FTA limit of detection was 7 µg/mL for A. invadans compared to 56 µg/mL for the immunodot. FTA and polymerase chain reaction (PCR) could detect A. invadans in fish tissue homogenates at a 10-11 dilution compared to a 10-8 dilution by immunodot. In fish suffering from natural cases of epizootic ulcerative syndrome (EUS) collected from Mangalore, India, FTA and PCR could detect A. invadans in 100% of the samples compared to 89.04% detected by immunodot. FTA reagents were stable and produced expected results for 4 months when stored at 4~8℃. This rapid test could serve as simple and cost-effective on-site screening tool to detect A. invadans in fish from EUS outbreak areas and in ports during the shipment of live or frozen fish.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.