To establish an efficient design process for centrifugal blood pumps, the results of computational fluid dynamics (CFD) analysis were compared to the results of flow visualization tests and hemolysis tests, using the Nikkiso centrifugal blood pump. CFD analysis revealed that the radial gap greatly affected the shear stress in the outlet diffuser. The hemolysis study also indicated a similar tendency. To see the flow behind the impeller, we conducted a comparative study between models with and without washout holes using the CFD technique. CFD analysis indicated that flow and pressure distributions behind the impeller were different between both models, and a particle was observed to remain longer behind the impeller in the model without washout holes. In the future, CFD analysis could be a useful tool for developing blood pumps in comparison to flow visualization tests and hemolysis tests.
There are few established engineering guidelines aimed at reducing hemolysis for the design of centrifugal blood pumps. In this study, a fluid dynamic approach was applied to investigate hemolysis in centrifugal pumps. Three different strategies were integrated to examine the relationship between hemolysis and flow patterns. Hemolytic performances were evaluated in in vitro tests and compared with the flow patterns analyzed by flow visualization and computational fluid dynamic (CFD). Then our group tried to establish engineering guidelines to reduce hemolysis in the development of centrifugal blood pumps. The commercially available Nikkiso centrifugal blood pump (HPM-15) was used as a standard, and the dimensions of 2 types of gaps between the impeller and the casing, the axial and the radial gap, were varied. Four impellers with different vane outlet angles were also prepared and tested. Representative results of the hemolysis tests were as follows: The axial gaps of 0.5, 1.0, and 1.5 mm resulted in normalized index of hemolysis (NIH) values of 0.0028, 0.0013 and 0.0008 g/100 L, respectively. The radial gaps of 0.5 and 1.5 mm resulted in NIH values of 0.0012 and 0.0008 g/100 L, respectively. The backward type vane and the standard one resulted in NIH values of 0.0013 and 0.0002 g/100 L, respectively. These results revealed that small gaps led to more hemolysis and that the backward type vane caused more hemolysis. Therefore, the design parameters of centrifugal blood pumps could affect their hemolytic performances. In flow visualization tests, vortices around the impeller outer tip and tongue region were observed, and their patterns varied with the dimensions of the gaps. CFD analysis also predicted high shear stress consistent with the results of the hemolysis tests. Further investigation of the regional flow patterns is needed to discuss the cause of the hemolysis in centrifugal blood pumps.
The purpose of this study was to examine the relationship between local flow conditions and the hemolysis level by integrating hemolysis tests, flow visualization, and computational fluid dynamics to establish practical design criteria for centrifugal blood pumps with lower levels of hemolysis. The Nikkiso centrifugal blood pump was used as a standard model, and pumps with different values of 3 geometrical parameters were tested. The studied parameters were the radial gap between the outer edge of the impeller vane and the casing wall, the position of the outlet port, and the discharge angle of the impeller vane. The effect of a narrow radial gap on hemolysis was consistent with no evidence that the outlet port position or the vane discharge angle affected blood trauma in so far as the Nikkiso centrifugal blood pump was concerned. The radial gap should be considered as a design parameter of a centrifugal blood pump to reduce blood trauma.
With a 250% scaled-up pump model, high speed video camera, and argon ion laser light sheet, flow patterns related to hemolysis were visualized and analyzed with 4 frame particle tracking software. Different flow patterns and shear distributions were clarified by flow visualization for pumps modified to have different hemolysis levels. A combination of in vitro hemolysis tests, flow visualization, and CFD analysis suggested a close relationship between hemolysis and high shear caused by small impeller/casing gaps. Because arbitrary cross sections can be illuminated by laser light sheet, flow visualization is a useful tool in finding locations related to hemolysis in the design process of rotary blood pumps.
To clarify the effect of washout holes on the flow in a centrifugal blood pump to prevent blood stagnation, a quantitative flow visualization technique was applied to compare flows in models with and without washout holes. A scaled-up model of a prototype pump and a high speed video camera were used for the flow visualization, and images were processed by particle tracking velocimetry. Particular attention was paid to the flow through the gaps behind and in front of the impeller. The results showed that in the gap behind the impeller, washout holes caused not only an inward flow, but also an increase in the tangential velocities. In the gap in front of the impeller, washout holes caused an outward flow and a decrease in the tangential velocities. Head flow characteristics were little affected by the washout holes in this initial design for which the flow through the washout holes was set to be approximately 10% of the flow in the external circuit. These results suggest that the flow through washout holes is significant in the prevention of blood stagnation in 2 ways. First, the inward radial velocity behind the impeller and outward velocity in front of the impeller result in fluid exchange, and second, a tangential velocity increase reduces fluid stagnation behind the impeller.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.