In 2013, we published a simple method for aortic valve reconstruction in children using fixed pericardium in this journal. This was not expected to be a durable solution but a temporizing measure in the absence of other solutions, in order to buy time for growth that would allow subsequent prosthetic valve insertion. As such, this was implemented in 3 patients with excellent immediate results, as reported in our original publication. We are now writing to provide longer term information. In 2 cases, a newborn and a 12-year-old boy, the valve calcified heavily at 4 years' follow-up. By then, the valvar annulus had grown sufficiently to accept a prosthetic valve in both cases. In these 2 cases, the material used for valve reconstruction was autologous pericardium fixed in 0.6% gluteraldehyde. In the third patient, a 3-year-old girl, the valve failed by dehiscence of the suture line 6 months following surgery. In this case, fixed bovine pericardium had been used, which is more rigid than autologous pericardium, thus possibly explaining the dehiscence. However, the pericardial patch was not calcified. Thus, the described method lived up to our expectation of buying time for growth for later insertion of a prosthetic valve in 2 of the 3 cases.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.