We designed a dedicated multi-detector multipinhole brain SPECT scanner to generate images of higher quality compared to general-purpose systems. The system, AdaptiSPECT-C, is intended to adapt its sensitivity-resolution trade-off by varying its aperture configurations allowing both high-sensitivity dynamic and high-spatial-resolution static imaging. The current system design consists of 23 detector heads arranged in a truncated spherical geometry. In this work, we investigated the axial and angular sampling capability of the current stationary system design. Two data acquisition schemes using limited rotation of the gantry and two others using axial translation of the imaging bed were also evaluated concerning their impact on image quality through improved sampling. Increasing both angular and axial sampling in the current prototype system resulted in quantitative improvements in image quality metrics and qualitative appearance of the images as determined in studies with specifically selected phantoms. Visual improvements for the brain phantoms with clinical distributions were less pronounced but presented quantitative improvements in the fidelity (normalized root-mean-square error (NRMSE)) and striatal specific binding ratio (SBR) for a dopamine transporter (DAT) distribution, and in NRMSE and activity recovery for a brain perfusion distribution. More pronounced improvements with increased sampling were seen in contrast recovery coefficient, bias, and coefficient of variation for a lesion in the brain perfusion distribution. The negligible impact of the most cranial ring of detectors on axial sampling, but its significant impact on sensitivity and angular sampling in the cranial portion of the imaging volume-of-interest were also determined. Index Terms-AdaptiSPECT-C, angular and axial sampling, SPECT, pinhole, multi-pinhole. I. INTRODUCTION B RAIN SPECT imaging has been of interest for many clinical applications in neuroimaging including Alzheimer's, Parkinson's, and Huntington's diseases,
With brain-dedicated multi-detector systems employing pinhole apertures the usage of detectors facing the top of the patient’s head (i.e. quasi-vertex (QV) views) can provide the advantage of additional viewing from close to the brain for improved detector coverage. In this paper, we report the results of simulation and reconstruction studies to investigate the impact of the QV views on the imaging performance of AdaptiSPECT-C, a brain-dedicated stationary SPECT system under development. In this design, both primary and scatter photons from regions located inferior to the brain can contribute to SPECT projections acquired by the QV views, and thus degrade AdaptiSPECT-C imaging performance. In this work, we determined the proportion, origin, and nature (i.e. primary, scatter, and multiple-scatter) of counts emitted from structures within the head and throughout the body contributing to projections from the different AdaptiSPECT-C detector rings, as well as from a true vertex view detector. We simulated phantoms used to assess different aspects of image quality (i.e. uniform activity concentration sphere, and Derenzo), as well as anthropomorphic phantoms with different count levels emulating clinical 123I activity distributions (i.e. DaTscan and perfusion). We determined that attenuation and scatter in the patient’s body greatly diminish the probability of the photons emitted outside the volume of interest reaching to detectors and being recorded within the 15% photopeak energy window. In addition, we demonstrated that the inclusion of the residual of such counts in the system acquisition does not degrade visual interpretation or quantitative analysis. The addition of the QV detectors improves volumetric sensitivity, angular sampling, and spatial resolution leading to significant enhancement in image quality, especially in the striato-thalamic and superior regions of the brain. Besides, the use of QV detectors improves the recovery of clinically relevant metrics such as the striatal binding ratio and mean activity in selected cerebral structures. Our findings proving the usefulness of the QV ring for brain imaging with 123I agents can be generalized to other commonly used SPECT imaging agents labelled with isotopes, such as 99mTc and likely 111In.
We are developing a multi-detector pinhole-based stationary brain-dedicated SPECT system: AdaptiSPECT-C. In this work, we introduced a new design prototype with multiple adaptable pinhole apertures for each detector to modulate the multiplexing by employing temporal shuttering of apertures. Temporal shuttering of apertures over the scan time provides the AdaptiSPECT-C with the capability of multiple-frame acquisition. We investigated, through analytic simulation, the impact of projection multiplexing on image quality using several digital phantoms and a customized anthropomorphic phantom emulating brain perfusion clinical distribution. The 105 pinholes in the collimator of the system were categorized into central, axial, and lateral apertures. We generated, through simulation, collimators of different multiplexing levels. Several data acquisition schemes were also created by changing the imaging time share of the acquisition frames. Sensitivity increased by 35% compared to the single-pinhole-per-detector base configuration of the AdaptiSPECT-C when using the central, axial, and lateral apertures with equal acquisition time shares within a triple-frame scheme with a high multiplexing scenario. Axial and angular sampling of the base configuration was enhanced by adding the axial and lateral apertures. We showed that the temporal shuttering of apertures can be exploited, trading the sensitivity, to modulate the multiplexing and to acquire a set of non-multiplexed non-truncated projections. Our results suggested that reconstruction benefited from utilizing both non-multiplexed projections and projections with modulated multiplexing resulting in a noticeably reduction in the multiplexing-induced image artefacts. Contrast recovery factor improved by 20% (9%) compared to the base configuration for a Defrise (hot-rod) phantom study when the central and axial (lateral) apertures with equal time shares were combined. The results revealed that, as an overall trend at each simulated multiplexing level, lowest normalized root-mean-square errors for the brain gray-matter regions were achieved with the combined usage of the central apertures and axial/lateral apertures.
Multi-pinhole (MPH) collimators are known to provide better trade-off between sensitivity and resolution for preclinical, as well as for smaller regions in clinical SPECT imaging compared to conventional collimators. In addition to this geometric advantage, MPH plates typically offer better stopping power for penetration than the conventional collimators, which is especially relevant for I-123 imaging. The I-123 emits a series of high-energy (>300 keV, ~2.5% abundance) gamma photons in addition to the primary emission (159 keV, 83% abundance). Despite their low abundance, high-energy photons penetrate through a low-energy parallel-hole (LEHR) collimator much more readily than the 159 keV photons, resulting in large downscatter in the photopeak window. In this work, we investigate the primary, scatter, and penetration characteristics of a single pinhole collimator that is commonly used for I-123 thyroid imaging and our two MPH collimators designed for I-123 DaTscan imaging for Parkinson’s Disease, in comparison to three different parallel-hole collimators through a series of experiments and Monte Carlo simulations. The simulations of a point source and a digital human phantom with DaTscan activity distribution showed that our MPH collimators provide superior count performance in terms of high primary counts, low penetration, and low scatter counts compared to the parallel-hole and single pinhole collimators. For example, total scatter, multiple scatter, and collimator penetration events for the LEHR were 2.5, 7.6 and 14 times more than that of MPH within the 15% photopeak window. The total scatter fraction for LEHR was 56% where the largest contribution came from the high-energy scatter from the back compartments (31%). For the same energy window, the total scatter for MPH was 21% with only 1% scatter from the back compartments. We therefore anticipate that using MPH collimators, higher quality reconstructions can be obtained in a substantially shorter acquisition time for I-123 DaTscan and thyroid imaging.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.