Background Lymphedema is a common complication of breast cancer treatment that affects one in five breast cancer survivors, yet there is no reliable method to detect lymphedema in the subclinical range. The objective of this study was to determine the feasibility and reliability of using an infrared 3D scanning device (ISD) as a peri-operative limb volume measurement tool. Methods Fifteen patients were analyzed based on inclusion criteria. Peri-operative measurements were obtained using tape measure and an ISD. Volumes were calculated using a standard algorithm for tape measure and a custom algorithm for ISD measurements. Linear regression models were used to assess ISD and tape measurement volume and circumference correlation. One-way ANOVA was used to compare change in percent difference at set time points post-operatively (2–3 weeks, 4–6 weeks, and 7–12 weeks) for both ISD and tape measure. t tests for unequal variances with the Bonferroni correction were performed among these groups. Results There is a positive linear correlation (R2 = 0.8518) between absolute volume measurements by the ISD and tape measure. Analyses over 2–10 weeks post-operatively showed that the ISD was able to detect volume changes in both the unaffected and the affected arm. Furthermore, the affected arm tended to have a greater increase in volume in the majority of patients, indicating these patients could be at risk for lymphedema. Conclusions Technology utilizing infrared 3D scanners can reliably measure limb volume pre- and post-treatment similarly to tape measure in a small sample of patients. Further research using 3D scanning technology with a longer follow up is warranted.
Background Lymphedema is a common complication of breast cancer treatment that affects one in five breast cancer survivors yet there is no reliable method to detect lymphedema in the subclinical range. The objective of this study was to determine the feasibility and reliability of using an infrared 3D scanning device (ISD) as a peri-operative limb volume measurement tool.Methods Fifteen patients were analyzed based on inclusion criteria. Peri-operative measurements were obtained using tape measure and an ISD. Volumes were calculated using a standard algorithm for tape measure and a custom algorithm for ISD measurements. Linear regression models were used to assess ISD and tape measurement volume and circumference correlation. One-way ANOVA was used to compare change in percent difference at set time points post-operatively (2–3 weeks, 4–6 weeks, and 7–12 weeks) for both ISD and tape measure. T-Tests for unequal variances with the Bonferroni correction were performed among these groups.Results There is a positive linear correlation (R-squared = 0.8518) between absolute volume measurements by the ISD and tape measure. Analyses over 2–10 weeks post-operatively showed that the ISD was able to detect volume changes in both the unaffected and the affected arm. Furthermore, the affected arm tended to have a greater increase in volume in the majority of patients, indicating these patients could be at risk for lymphedema.Conclusions Technology utilizing infrared 3D scanners can reliably measure limb volume pre and post treatment similarly to tape measure in a small sample of patients. Further large-scale studies are needed to validate this method in a diverse population.
Background Lymphedema is a common complication of breast cancer treatment that affects one in five breast cancer survivors yet there is no reliable method to detect lymphedema in the subclinical range. The objective of this study was to determine the feasibility and reliability of using an infrared 3D scanning device (ISD) as a peri-operative limb volume measurement tool.Methods Fifteen patients were analyzed based on inclusion criteria. Peri-operative measurements were obtained using tape measure and an ISD. Volumes were calculated using a standard algorithm for tape measure and a custom algorithm for ISD measurements. Linear regression models were used to assess ISD and tape measurement volume and circumference correlation. One-way ANOVA was used to compare change in percent difference at set time points post-operatively (2-3 weeks, 4-6 weeks, and 7-12 weeks) for both ISD and tape measure. T-Tests for unequal variances with the Bonferroni correction were performed among these groups.Results There is a positive linear correlation (R-squared = 0.8518) between absolute volume measurements by the ISD and tape measure. Analyses over 2-10 weeks post-operatively showed that the ISD was able to detect volume changes in both the unaffected and the affected arm. Furthermore, the affected arm tended to have a greater increase in volume in the majority of patients, indicating these patients could be at risk for lymphedema. Conclusions Technology utilizing infrared 3D scanners can reliably measure limb volume pre and post treatment similarly to tape measure in a small sample of patients. Further research using 3D scanning technology with a longer follow up is warranted.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.