A procedure for dynamic force calibration using sinusoidal excitations of force transducers is described. The method is based on a sinusoidal excitation of force transducers equipped with an additional top mass excited with an electrodynamic shaker system. The acting dynamic force can in this way be determined according to Newton's law as mass times acceleration, whereby the acceleration is measured on the surface of the top mass with the aid of laser interferometers. The dynamic sensitivity, which is the ratio of the electrical output signal of the force transducer and the acting dynamic force, is the main point of interest of such a dynamic calibration. In addition to the sensitivity, the parameter stiffness and damping of the transducer can also be determined. The first part of the paper outlines a mathematical model to describe the dynamic behaviour of a transducer. This is followed by a presentation of the traceability of the measured quantities involved and their uncertainties. The paper finishes with an example calibration of a 25 kN strain gauge force transducer.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.