Large skull defects as a result of craniectomies due to cerebral insults, trauma, or tumors create functional and aesthetic disturbances for the patient. Cranioplasty with implants in these cases are an alternative to autogenous bone transplantation. In our clinic, customized titanium or optima poly-ether-ether ketone (PEEK) implants are used to reconstruct craniectomy defects. To compare the two materials we investigated the structural changes of the implants fixed to a sintered polyamide skull model under mechanical stress in four simplified models. In a standard testing machine, the models were subjected to a load under a quasi-static loading rate of 1.925 mm/min. Fractures of the PEEK implants occurred at a force of 24.2 and 24.5 kN with a displacement of 8.4 and 8 mm. The titanium implants showed no deformation, but extensive damage was seen in the polyamide skull models. The highest pressures achieved were 45.8 and 50.9 kN. In a simplified model with quasi-static loading, both implants withstood forces that were higher than those capable of causing skull fractures. It seems that the mechanical properties of PEEK could provide better protection when used for cranioplasty in patients after craniectomy if reconstruction with autogenous bone is not possible.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.