Background:
Shape-memory abnormalities are seen in some nitinol atrial septal occluders. Variably described as cobra-head, tulip, and others, their incidence, mechanisms, clinical impact, and outcome have not been systematically analysed.
Methods:
We retrospectively reviewed all consecutive device closures in the last 6 years for deformations. Type and size of the occluder, deployment technique, size, and angulation/kinking of the delivery sheath were analysed. Procedural success, duration, and other complications were studied.
Results:
A total of 112 devices (11.8%) among 950 occluders used in 936 patients showed deformities. Fourteen of 936 received 2 devices. Deformities were transient and self-correcting in 40%. Multivariate analysis showed significant associations with oversized sheaths (p = 0.004), kinked/angulated sheaths (p < 0.001), special deployment techniques (p < 0.001), and twist in the device waist (p = 0.011). Despite more frequent deformities with Figulla (15.6%) and Amplatzer (13.9%) occluders than Cera occluders (6.6%) and larger devices (>24 mm – 14.6%) than smaller devices (less than or equal to 24 mm – 9.7%), they were not significant on multivariate analysis. In vivo manipulations corrected most deformities; nineteen needed in vitro reformations and four needed a change of device. Despite prolongation of the procedure, repeated attempts (mean 2.76 ± 1.7 attempts, with a range from 1 to 9 attempts), and supraventricular tachycardia in two patients, there were no serious adverse effects.
Conclusions:
Deformations were frequent in 11.8% of atrial septal occluders on a targeted search. Oversized and angulated/kinked sheaths, special techniques like pulmonary vein deployment and twist in device waist during procedure predisposed to deformities. While most deformities were corrected with manipulations, removal of the device was infrequently needed and change of device was rarely required. Long procedural time and multiple attempts for deployment did not affect procedural success.
Coral reef aorta is a rare calcifying obstructive disease that involves the thoracoabdominal aorta. Similar presentations in the postsubclavian aorta may result in acquired atheromatous aortic coarctation leading to systemic hypertension and heart failure. The associated calcification makes surgical anatomic or extraanatomic bypass and thromboendarterectomy challenging. Extensive circumferential calcification often precludes endovascular intervention. We present the case of a 25-year-old man with an acquired atheromatous coarctation of the postsubclavian aorta who underwent successful endovascular treatment with use of a balloon-expandable covered stent.
A diagnostic coronary catheter injury to the subaortic region in a 41-year-old woman with rheumatic heart disease led to a pseudoaneurysm that later caused extrinsic left coronary compression. She subsequently underwent double-valve replacement, overlooking the pseudoaneurysm that enlarged to a giant size three months later following thrombolysis for mitral prosthesis thrombosis. A thrombolysis-induced large intracerebral hemorrhage posed a significant risk for reoperation, and mechanical prosthetic valves in the aortic and mitral positions allowed a catheter option only, through percutaneous transapical access. Interventional closure of the pseudoaneurysm is discussed in this unique report.
Fontan surgery streamlines the systemic venous return through the pulmonary circulation before filling the systemic ventricle in univentricular hearts. The venous congestion leads to effusions, lowers cardiac output, and affects organ perfusion. Fenestrations in the Fontan circuit improve forward flow through the ventricles, lower venous pressures, and reduce perioperative morbidity. When large fenestrations cause profound hypoxia and effort intolerance, there are no current techniques to reduce their size. Atrial flow regulators with a predetermined orifice were used off-label in three borderline patients with large undesirable fenestrations following extracardiac conduit Fontan surgeries. This resulted in improved oxygenation without marked elevation of venous pressures, while retaining the patency of the decompressive fenestration.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.