Cuff pressure gauges are the only recommended instrument to perform controls on endotracheal tube cuff pressure during anesthesia. No calibration is mandatory for these devices. The aim of this study was to describe the level of conformity of various cuff pressure gauges. The single-center measurements were performed with a cuff pressure calibrator on all cuff pressure gauges that were usually used in the operating room. Seven measurements (repeat three times) on each cuff pressure gauges at different levels of pressures (i.e. 0, 10, 20, 27, 30, 40 and 50 cmHO) were performed. Our homologation criteria were either the reliability of the leak test (value of cuff pressure gauges maintained at 120 cmHO during 5 s) or the difference between the values of the cuff pressure tested and the calibrator below 1.3 cmHO at the range of 20-30 cmHO. A total of 567 measurements on 27 cuff pressure gauges were performed. Only 30% (n = 8/27) of the cuff pressure gauges reach our homologation criteria. 30% (n = 8/27) failed at the leak test. 48% (n = 13/27) of the cuff pressure gauges tested, had a calibration variation error > 1.3 cmHO on the levels of pressure between 20 and 30 cmHO. A minority of cuff pressure gauges went through our homologation criteria. These results demonstrate us that there is a real problem of the reliability and the follow-up of those medical devices. This study suggests to reinforce biomedical engineering control on these devices.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.