A method is described to measure the water activity (aw) of foods in the range of 0.935 to 1.0 using a psychrometer. Samples (0.15 to 0.35 g) were equilibrated for 2 h in the sample chamber before cooling the thermocouple and recording the subsequent voltage on the instrument. The aw was read off a standard curve prepared from sodium chloride solutions of known water activity values. Results show that the method was considerably better than the graphical interpolation method and comparable to the microcrystalline method but was much quicker and required less effort.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.