Accurately measuring Rn-220 is crucial for evaluating natural radiation exposure. For the large volume of normal scintillation cell, the detection efficiencies for different energy of α particles decayed from Rn-222, Rn-220 and their progeny are varied. To address this, two systems of the low-pressure scintillation cell were designed to make the shortest range of α particles with minimum energy decayed from Rn-222 longer than the longest distance between any two points in the scintillation cell by reducing air pressure. Thus, the detection efficiencies of the low-pressure scintillation cell for α particles decayed from Rn-222, Rn-220 and their progeny are same. The detection efficiency of low-pressure scintillation cell for α particles can be obtained by the standard radon chamber, and then the calibration factor of Rn-220 can be determined by low-pressure scintillation. Several verification experiments were performed. The experimental results show that the detection efficiencies of Rn-222 and Rn-220 at the air pressure of 0.4P are very close and become saturated for lower values of air pressure. The calibration factor of Rn-220 calculated using the detection efficiency (0.76 ± 0.04) at the air pressure of 0.4P is 53.75 ± 2.96 Bq m-3 min-1 (k=1).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.