Due to the strong level of evidence of excellent test-retest reliability that supports use of the Nana positioning protocol, it is recommended as the first choice for clinicians when using DXA to assess body composition.
The Nana positioning protocol is widely used to position participants to minimize technical error when undertaking body composition scanning and analysis with a Dual energy X-Ray absorptiometry (DXA) machine. Once biological and technical errors are accounted for, the only variation in test-retest results is from statistical fluctuation or machine error. Therefore, the aim of this study is to assess the test-retest reliability of the Nana positioning protocol and establish the smallest real difference percentage (SRD%). A gender-balanced group of 30 participants (15 males, 15 females) underwent two scans in succession using the Nana positioning protocol, with repositioning between scans. Percentage change in mean with typical error, Intraclass Correlation Coefficients (ICC), and standard error measurement percentage (SEM%) were used to identify the test-retest reliability and error rate of these protocols. Additionally, SRD% was calculated to assess the point at which clinically important changes occurred in a participant. The reliabilities of the whole body and regional scans were excellent. Percentage change in mean ranged between 0.00% and 0.23%. High reproducibility of the Nana positioning protocol was evident through an ICC ranging between 0.966-1.000. Additionally, typical error, SEM%, and SRD% were all low. Interestingly, fat mass was associated with the largest fluctuations observed to be associated with any of the parameters assessed. When all sources of biological and technical errors have been accounted for, the Nana positioning protocol has excellent test-retest reliability and produces low SEM% and SRD%.
1 1 2 3 4 Does a novice technician produce results similar to that of an experienced DXA 5 technician when assessing body composition and bone mineral density? 6 7 8
BackgroundDual energy X-ray absorptiometry (DXA) is a commonly used instrument for analysing segmental body composition (BC). The information from the scan guides the clinician in the treatment of conditions such as obesity and can be used to monitor recovery of lean mass following injury. Two commonly used DXA positioning protocols have been identified—the Nana positioning protocol and the National Health and Nutrition Examination Survey (NHANES). Both protocols have been shown to be reliable. However, only one study has assessed the level of agreement between the protocols and ascertained the participants’ preference of protocol based upon comfort. Given the paucity of research in the field and the growing use of DXA in both healthy and pathological populations further research determining the most appropriate positioning protocol is warranted. Therefore, the aims of this study were to assess the level of agreement between results from the NHANES protocol and Nana protocol, and the participants’ preference of protocol based on comfort.MethodsThirty healthy participants (15 males, 15 females, aged 23–59 years) volunteered to participate in this study. These participants underwent two whole body DXA scans in a single morning (Nana positioning protocol and NHANES positioning protocol), in a randomised order. Each participant attended for scanning wearing minimal clothing and having fasted overnight, refrained from exercise in the past 24 h and voided their bladders. Level of agreement, comparing NAHNES to Nana protocol was assessed using an intra-class correlation coefficient (ICC), concordance correlation coefficient (CCC) and percentage change in mean. Limit of agreement comparing the two protocols were assessed using plots, mean difference and confidence limits. Participants were asked to indicate the protocol they found most comfortable.ResultsWhen assessing level of agreement between protocols both the ICC and CCC scores were very high and ranged from 0.987 to 0.997 for whole body composition, indicating excellent agreement between the Nana and NHANES protocols. Regional analysis (arms, legs, trunk) ICC scores, ranged between 0.966 and 0.996, CCC ranged between 0.964 and 0.997, change in mean percentage ranged between −0.58% and 0.37% which indicated a very high level of agreement. Limit of agreement analysis using mean difference ranged between −0.223 and 0.686 kg and 95% CL produced results ranging between −1.262 kg and 1.630 kg. The majority (80%) of participants found the NHANES positioning protocol more comfortable.DiscussionThis study reveals a strong level of agreement as illustrated by high ICC’s and CCC’s between the positioning protocols, however systematic bias within limit of agreement plot and a large difference in 95% confidence limits indicates that the protocols should not be interchanged when assessing an individual. The NHANES protocol affords greater participant comfort.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.