This paper aims to present the unique, to the best of our knowledge, case of entrapment of a standard vacuum drainage tube in the articulating surfaces of the cup of dual-mobility total hip arthroplasty. A 75year-old woman with end-stage idiopathic avascular necrosis of the left femoral head was referred to the arthroplasty service of our tertiary orthopedic department. She underwent a scheduled and uneventful total hip arthroplasty with a press-fit dual-mobility prosthesis through a standard posterior approach. On the second postoperative day, the attempt to remove the standard vacuum drainage was unsuccessful. Consequently, the patient underwent urgent re-operation. The drain tube was found entrapped between the articulating surfaces of the posterior-inferior aspect of the dual-mobility cup and was uneventfully removed. The patient was discharged with no further events three days after her second operation. Our unique rare case increases awareness when performing even routine everyday surgical procedures because a rare complication may occur irrespective of the level of vigilance of the surgeon and can potentially compromise the outcomes of an otherwise well-performed operation.
No abstract
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.