This Technology Overview was prepared using systematic review methodology and summarizes the findings of studies published as of July 15, 2011, on modern metal-on-metal hip implants. Analyses conducted on outcomes by two joint registries indicate that patients who receive metal-on-metal total hip arthroplasty (THA) and hip resurfacing are at greater risk for revision than are patients who receive THA using a different bearing surface combination. Data from these registries also indicate that larger femoral head components have higher revision rates and risk of revision and that older age is associated with increased revision risks of large-head metal-on-metal THA. Several studies noted a correlation between suboptimal hip implant positioning and higher wear rates, local metal debris release, and consequent local tissue reactions to metal debris. In addition, several studies reported elevated serum metal ion concentrations in patients with metal-on-metal hip articulations, although the clinical significance of these elevated ion concentrations remains unknown.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.