Cardiopulmonary resuscitation (CPR) is alongside electrical defibrillation the most crucial countermeasure for sudden cardiac arrest, which affects thousands of individuals every year. In this paper, we present a novel approach including sinusoid models that use skeletal motion data from an RGB-D (Kinect) sensor and the Differential Evolution (DE) optimization algorithm to dynamically fit sinusoidal curves to derive frequency and depth parameters for cardiopulmonary resuscitation training. It is intended to be part of a robust and easy-to-use feedback system for CPR training, allowing its use for unsupervised training. The accuracy of this DE-based approach is evaluated in comparison with data of 28 participants recorded by a state-of-the-art training mannequin. We optimized the DE algorithm hyperparameters and showed that with these optimized parameters the frequency of the CPR is recognized with a median error of ±2.9 compressions per minute compared to the reference training mannequin.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.