The AAMI working group ST/WG 93 is finalizing a standard (AAMI ST98) for the cleaning validation of reusable medical devices based on guidance from the technical information report AAMI TIR30:2011/(R)2016. A number of analytical best practices are being considered for this new standard. Test method suitability for processing cleaning validations historically has been established using one positive control and performing an extraction efficiency. The new cleaning validation standard is proposed to require a change from only one replicate test sample to three when performing method suitability. This change will affect manufacturers; therefore, the value of and consideration for performing these additional replicates requires explanation. This article discusses how variation of validation parameters can affect the accuracy and precision during method suitability testing. Multiple replicates are needed to understand the variability of method extraction and impact on cleaning validations of reusable medical devices.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.