Abstract. Nine european national metrology institutes (NMIs) are collaborating in a new project funded by the european metrology research programme (EMRP) to establish traceable dynamic measurement of the mechanical quantities force, pressure, and torque. The aim of this joint research project (JRP) is to develop appropriate calibration methods, mathematical models, and uncertainty evaluation. The duration of the project is 3 years for a global amount of e3.6 million. It began in September 2011.
This is the final report for CIPM key comparison CCAUV.V-K3 in the area of 'vibration' (quantity of acceleration). The aim of this comparison was to measure the voltage sensitivity of one accelerometer standard set with primary means at 27 frequencies from 0.1 Hz to 40 Hz. Fourteen Metrology Institutes from five RMOs have participated in the comparison with National Institute of Metrology, P.R. China as pilot lab and Laboratoire National de Métrologie et d'Essais and National Metrology Institute of South Africa as co-pilot labs. One quartz-flexure servo accelerometer of single-ended type and a signal conditioner was circulated among the participants. All but one of the participating laboratories provided their calibration results, which were mostly consistent within their declared expanded uncertainties for magnitude results. Only two participants failed to contribute to the KCRV values calculated for five frequencies. For phase shift, three participants could not contribute to the calculation of the KCRV values in a total of sixteen frequencies. This first low-frequency vibration key comparison revealed the current calibration capabilities of the fourteen participants of five RMOs.
Main text
To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Accelerometer calibration by primary and secondary means is now very well covered both in low and medium frequency range. This is obviously not the case for seismometers which are widely used in the world. Up to now, seismometers only have internal checks thanks to an integrated coil. Calibration of these devices address different challenges like large mass and volumes, non-rotatable devices, specific mounting. This inter laboratory comparison will help to investigate these fields and to give feedback to develop a methodology to calibrate seismometers.
The results of a Pilot Study EURAMET.AUV.V-P1 between LNE (France) and RCM-LIPI (Indonesia) are reported. This bilateral comparison of sinusoidal vibration was organized after the implementation of various improvements at RCM-LIPI following a previous (unpublished) comparison that had revealed discrepancies in their results at frequencies above 5 kHz.
The results of this Pilot Study, using the same accelerometers as the earlier comparison, demonstrate that the discrepancies at high frequencies have been resolved. For both the back-to-back and the single-ended accelerometers tested, the sensitivities of the RCM-LIPI and the LNE over the frequency range from 10 Hz to 10 kHz now agree within their declared uncertainties.
Main text
To reach the main text of this paper, click on Final Report.
The final report has been peer-reviewed and approved for publication by the CCAUV.
Abstract. In order to protect from occupational disease related to vibrations, the European Directive2002/44/CE defines alert and action limits which trigger an action on the vibrations received by the workers. Since the 6th July 2010, companies have been compelled to assess vibration doses thanks to a vibration dosimeter.LNE has been willing to develop its activities in accelerometry, LNE developed the calibration of vibration dosimeter service following standard ISO 8041 "Human Response to vibration / measuring instrumentation".This article describes the requirements of the European Directive 2002/44/CE, the standard ISO 8041 and the calibration bench developed and accredited by COFRAC.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.