Abstract. MobiusFX, an add-on software module from Mobius Medical Systems for IMRT and VMAT QA, uses measurements in linac treatment logs to calculate and verify the 3D dose delivered to patients. In this study, 10 volumetric-modulated arc therapy (VMAT) prostate plans were planned and delivered in a Varian TrueBeam linac. The plans consisted of beams with 6 and 10 MV energy and 2 to 3 arcs per plan. The average gamma value with criterion of 3% and 3mm MobiusFX and TPS: 99.96%, 2% and 2mm MobiusFX and TPS: 98.70 %. Further comparison with ArcCheck measurements was conducted.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.