The ICare tonometer can be useful in a routine clinical setting. The IOP readings are quite in accordance with those obtained by GAT. The measurements seemed to be influenced by CCT variations, and thus pachymetry should always be taken into consideration.
ABSTRACT.Purpose: To compare intraocular pressure (IOP) measurements taken with Pascal dynamic contour tonometry (DCT), the TonoPen and the Goldmann applanation tonometry (GAT). The influence of central corneal thickness (CCT) on IOP measurements taken with Pascal DCT and the TonoPen was evaluated.Methods: One eye in each of 101 consecutive patients with primary open-angle glaucoma (POAG) underwent ultrasonic CCT measurement and IOP evaluation with GAT, Pascal DCT and the TonoPen in random order. The agreement between results from Pascal DCT and the TonoPen and those of GAT was assessed using the Bland)Altman method. The deviation of Pascal DCT and TonoPen readings from GAT values, corrected for CCT, was calculated and correlated to CCT using a linear regression model.
Results:The mean of the differences in IOP measurements was 3.2 ± 2.4 mmHg for Pascal DCT minus GAT readings and 0.5 ± 4.5 mmHg for TonoPen minus GAT readings. The 95% confidence interval of differences in IOP measurements was higher between TonoPen and GAT readings () 6 to 7 mmHg) than between Pascal and GAT readings (0.1)6.8 mmHg). Pascal DCT significantly overestimated IOP compared with GAT, especially for higher IOP readings. Bland)Altman scatterplots showed reasonable inter-method agreement between Pascal DCT and GAT measurements, and poor agreement between TonoPen and GAT measurements. The deviations of Pascal DCT and TonoPen readings from the corrected GAT values were both highly correlated with CCT values (linear regression analysis, p < 0.0001). The mean change in measured IOP for a 10-lm increase in CCT was 0.48 mmHg for Pascal DCT and 0.74 mmHg for the TonoPen.Conclusions: Agreement with GAT measurements was higher for Pascal DCT than for TonoPen readings; however, Pascal DCT significantly overestimated IOP values compared with GAT. Measurements of IOP obtained with both Pascal DCT and the TonoPen appeared to be influenced by CCT, and this influence appeared to be greater for the latter.
Corvis-ST precision was excellent for IOP and CCT; moderate for CDPs. Corvis-ST underestimated GAT IOP, especially at higher IOP and at lower corneal deformability levels. GAT and Corvis-ST IOP increased in thicker and less deformable corneas. Glaucoma patients showed significantly less deformable corneas than controls.
Purpose To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT) and iCare tonometry in normal and post-keratoplasty corneas and to assess the influence of central corneal thickness (CCT), corneal curvature (CC), and corneal astigmatism (CA) on IOP. Methods This prospective cross-sectional study included one eye of 101 subjects with normal corneas (58 healthy subjects, 43 glaucoma); and 90 post-keratoplasty patients: 34 penetrating keratoplasties (PK); 20 automated-lamellar-therapeutic keratoplasties (ALTK); 19 Descemet-stripping-automatedendothelial keratoplasties (DSAEK); 17 edematous grafts. All subjects underwent GAT and iCare IOP measurements in random order, and CCT, CC, and CA evaluation. The Bland-Altman method and multivariate regression analysis were used to assess inter-tonometer agreement and the influence of CCT, CC, and CA on IOP.Results iCare significantly underestimated IOP in all groups compared with GAT (GAT minus iCare of 3.5±3.5 mm Hg, Po0.001), but overestimated IOP in the edematous grafts (GAT minus iCare of À6.5 ± 1.9 mm Hg, Po0.001). In normal corneas, both tonometer measurements were directly related to CCT values; iCare readings appeared inversely related to CC. There was no significant relationship between IOP and CCT, CC and CA in post-keratoplasty eyes, except between CC and iCare measurements for PK eyes. Conclusions The agreement between GAT and iCare was clinically acceptable in control, ALTK and DSAEK groups, and poor in PK and edematous grafts eyes. In normal corneas, GAT was significantly affected by CCT; iCare was influenced by CCT and CC. The iCare appeared less influenced by corneal edema when compared with GAT. High IOP readings taken with both tonometers in grafts should raise suspicion of true elevated IOP.
ABSTRACT.Purpose: To compare the abilities of scanning laser polarimetry with the variable corneal compensator (GDx VCC) with those of optical coherence tomography (Stratus OCT) in discriminating between healthy and early-to-moderate perimetric glaucomatous eyes. Methods: A total of 95 glaucomatous patients (mean deviation ) 3.7 ± 3.0 dB, pattern standard deviation 4.5 ± 2.7 dB) and 62 control subjects underwent imaging by the GDx VCC and Stratus OCT using both optic nerve head (ONH) and retinal nerve fibre layer (RNFL) scan protocols. One eye per patient was considered. Sensitivity at ‡ 90% specificity and area under the receiver operating characteristic curve (AROC) were calculated for each GDx VCC and Stratus OCT index. Results: The largest AROCs with Stratus OCT were associated with cup : disc area ratio (0.88) for ONH scan indices, and with average thickness (0.84) for RNFL scan indices. The nerve fibre indicator provided the greatest AROC for the GDx VCC indices (0.85). Conclusions: Both the GDx VCC and Stratus OCT instruments were shown to be useful in the detection of glaucomatous damage. The best performing indices for the GDx VCC and Stratus OCT with both ONH and RNFL scans gave similar AROCs, showing a moderate sensitivity in early-to-moderate glaucoma patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.