A newly developed Digital Radiography (DR) detector has smaller pixel size and higher fill factor than earlier detector models. These technical advantages should theoretically lead to higher sensitivity and higher spatial resolution, thus making dose reduction possible without scarifying image quality compared to previous DR detector versions. To examine whether the newly developed Canon CXDI-70C DR detector provides an improved image quality and/or allows for dose reductions in hand and pelvic bone examinations as well as premature chest examinations, compared to the previous (CXDI-55C) DR detector version. A total of 450 images of a technical Contrast-Detail phantom were imaged on a DR system employing various kVp and mAs settings, providing an objective image quality assessment. In addition, 450 images of anthropomorphic phantoms were taken and analyzed by three specialized radiologists using Visual Grading Analysis (VGA). The results from the technical phantom studies showed that the image quality expressed as IQF INV values was on average approximately 45 % higher with the CXDI-70C detector compared to the CXDI-55C detector. Consistently, the VGA results from the anatomical phantom studies indicated that by using the CXDI-70C detector, diagnostic image quality could be maintained at a dose reduction of in average 30 %, depending on anatomy and kVp level. This indicates that the CXDI-70C detector is significantly more sensitive than the previous model, and supports a better clinical image quality. By using the newly developed DR detector a significant dose reduction is possible while maintaining image quality.
The purpose was to examine if scatter correction software could replace a grid while maintaining image quality and reducing radiation dose for pelvic DR examinations. Grid images was produced with 70 kV and 16mAs. Anthropomorphic- and Contrast Detail RADiography (CDRAD) non-grid images were produced with 60 kV, 80 kV and 90 kV combined with five different mAs and scatter correction software. The anthropomorphic images were analyzed by absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed using the CDRAD analysis software. The results showed a total of 54.6% non-grid images were evaluated as unsuitable for diagnostic use by the VGA. The CDRAD grid images showed that the IQF_inv values were significantly different (p = 0.0001) when compared to every group of non-grid images. Hereby, the conclusion stated that the scatter correction software did not compensate for the loss in image quality due to scattered radiation at the exposure levels included in a pelvic examination.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.