Modular acetabular systems have several advantages; however, increased polyethylene wear has been recognized as a disadvantage. In our study, an acetabular shell design with a highly polished inner surface, a locking mechanism that minimizes micromotion, and a high polyethylene liner to shell conformity was evaluated. A prospective cohort of 50 consecutive hips was followed up for a mean of 3.6 years (range, 2-7 years). One patient required an early revision for recurrent instability. Mean linear head penetration rates were 0.242 mm/year (range, 0.048-0.655 mm/year). The true linear wear rate was calculated after subtracting the linear wear in the first follow-up year from the total wear at the end of the follow-up. The true linear wear rate was 0.173 mm/year (range, 0.03-0.423 mm/year). A positive correlation was found between wear rate and male gender. The low head penetration rates suggest that a polished liner with an improved locking mechanism and increased liner conformity can reduce polyethylene wear.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.