The purpose is to devise a patient‐specific quality assurance procedure for RapidArc radiotherapy using the MapCHECK detector array. We use our existing MapCHECK system and a Solid Water phantom with an embedded ion chamber to develop a quality assurance procedure for RapidArc treatment after commissioning. The ion chamber used to measure the absolute dose is surrounded by 6 cm layers of solid water on the anterior and posterior sides. Partial arcs derived from the treatment planning system were used with MapCHECK to determine the actual shape of the dose and correct for the angular dependence. The ion chamber measurements were within 1% of the absolute doses predicted by the Eclipse treatment system. When using a partial arc from 60° to 300° on the MapCHECK array (gamma index <1: 3%, 3 mm, 10% threshold), we obtain a 97.52% average passing rate. A combination of ion chamber phantoms, partial arcs and the MapCHECK system can be used for quality assurance of RapidArc therapies.PACS number: 3.6.96.0
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.