A new cascade impactor has been designed specifically for pharmaceutical inhaler testing. This impactor, called the Next Generation Pharmaceutical Impactor (NGI), has seven stages and is intended to operate at any inlet flow rate between 30 and 100 L/min. It spans a cut size (D50) range from 0.54-microm to 11.7-microm aerodynamic diameter at 30 L/min and 0.24 microm to 6.12 microm at 100 L/min. The aerodynamics of the impactor follow established scientific principles, giving confident particle size fractionation behavior over the design flow range. The NGI has several features to enhance its utility for inhaler testing. One such feature is that particles are deposited on collection cups that are held in a tray. This tray is removed from the impactor as a single unit, facilitating quick sample turn-around times if multiple trays are used. For accomplishing drug recovery, the user can add up to approximately 40 mL of an appropriate solvent directly to the cups. Another unique feature is a micro-orifice collector (MOC) that captures in a collection cup extremely small particles normally collected on the final filter in other impactors. The particles captured in the MOC cup can be analyzed in the same manner as the particles collected in the other impactor stage cups. The user-friendly features and the aerodynamic design principles together provide an impactor well suited to the needs of the inhaler testing community.
An extension of the archival calibration of the recently developed 30-100-L/min seven-stage impactor, the Next Generation Pharmaceutical Impactor (NGI), has been undertaken at 15 L/min. The NGI stage cut sizes are 0.98-14.1 microm aerodynamic diameter at this flow rate. This 15-L/min calibration was motivated by the desire to sample the entire aerosol produced by a nebulizer when tested in accordance with a new international standard developed by the Comite Européen de Normalisation (CEN), as well as the need to test various types of inhalers at flow rates lower than 30 L/min for pediatric applications. Measurements were undertaken with monodisperse oleic acid droplets in the range of 0.7-22 microm aerodynamic diameter following a procedure established in the original 30-100-L/min calibration study. The NGI was found to be effective for particle size separation at 15 L/min. Users should decide the most applicable configuration that meets their needs, based on the following recommendations: (1) the pre-separator should not normally be used, as its performance is significantly degraded by the influence of gravity, resulting in interference with stage 1; and (2) a filter should be inserted below the micro-orifice collector (MOC), as the size corresponding to 80% collection efficiency of the MOC becomes excessively large with decreasing flow rate, so that this component becomes ineffective as a means of collecting fine particles that penetrate beyond stage 7.
A particle sampler has been developed, built, and tested. The sampler draws ambient air at approximately 300 L/min and then splits the sampled air into a particle-rich sample stream and a particle-depleted, reject stream.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.