In-vivo porcine studies on the effect of nanosecond high voltage pulses on liver tissue have shown that cell death can be induced in well-defined tissue volumes without damaging collagen-predominant structures. Comparison of the experimental results with the results of a three-dimensional finite element model allowed us to determine the threshold electric field for cell death. For 30, 100 nanosecond long pulses this was found to be in the range from 12 to 15 kV/cm. Modelling of the temperature distribution in the tissue using Pennes' bioheat equation showed that the lethal effect of nanosecond pulses on cells is non-thermal. Muscle contractions, generally caused by high voltage pulses, were significantly reduced for the 100 nanosecond pulses compared to microsecond long pulses. The results of these studies indicate that high voltage nanosecond pulses reliably kill normal liver cells in vivo and therefore may be useful for liver tumor treatments.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.