Abstract-A life-size mechanical middle ear model and human temporal bones were used to evaluate three different middle ear transducers for implantable hearing aids: the driving rod transducer (DRT), the floating mass transducer (FMT) or vibrant sound bridge, and the contactless transducer (CLT). Results of the experiments with the mechanical model were within the range of the results for human temporal bones. However, results with the mechanical model showed better reproducibility. The handling of the mechanical model was considerably simpler and less time-consuming. Systematic variations of mounting parameters showed that the angle of the rod has virtually no effect on the output of the DRT, the mass loading on the cable of the FMT has a larger impact on the output than does the tightness of crimping, and the output level of the CLT can be increased by 10 dB by optimizing the mounting parameters.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.