Due to increased experience and advanced techniques, percutaneous tracheostomy has become a popular, relatively safe procedure in the intensive care unit. The question of appropriate timing, however, has not been definitely answered with a randomized controlled trial. Instead, a number of retrospective studies and a single prospective study have shed some light on this issue. Most reports favor the performance of tracheostomy within 10 days of respiratory failure.
BACKGROUND:
The management of perioperative bleeding and the optimization of the available therapies are subjects of significant clinical interest. Clinical guidelines recommend the use of whole blood viscoelastic testing devices to target the utilization of blood products during major surgical procedures. The Quantra QPlus System is a new cartridge-based viscoelastic testing device based on an innovative ultrasound technology. The aim of this study was to evaluate this new system in a surgical population.
METHODS:
Two hundred seventy-seven adult subjects were enrolled in a multicenter, prospective observational study consisting primarily of patients undergoing cardiac and major orthopedic surgeries. Samples were obtained at multiple time points for testing on the Quantra QPlus System, the rotational thromboelastometry (ROTEM) delta, and standard coagulation tests. Quantra measurements included Clot Time (CT), Heparinase Clot Time (CTH), Clot Time Ratio (CTR), Clot Stiffness (CS), Fibrinogen (FCS), and Platelet (PCS) Contributions to CS. Data analyses included assessment of the concordance of Quantra parameters with a series of clinical composite indexes formed on the basis of standard coagulation tests in 3 domains representing increased, decreased, and normal/subclinical coagulation function. Linear regression and receiver operator characteristic (ROC) analyses of Quantra parameters with corresponding parameters from ROTEM assays were also performed.
RESULTS:
The accuracy (overall percent agreement or ratio of true positives and true negatives over the entire population) between the Quantra and the composite indexes was between 72% and 98% depending on the specific parameter. Linear regression analysis indicated that the correlation between ROTEM delta and Quantra was very strong with r values ranging between 0.84 and 0.89. Results from ROC analysis demonstrated sensitivities and specificities in the 80%–90% range when QPlus parameters were used to discriminate ROTEM threshold values currently used in goal-directed treatment algorithms.
CONCLUSIONS:
This study demonstrated that the Quantra QPlus System is strongly correlated with a well-established viscoelastic testing device and its parameters effectively represent the results from multiple standard laboratory assays. The Quantra has been designed to operate at the point of care with the potential to provide rapid and comprehensive results to aid in the management of coagulopathic patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.