The purpose of this study was to test a newly developed decontamination and fluidization kit for processing respiratory specimens for the detection of mycobacteria: the Myco-TB procedure (developed by Copan (Brescia, Italy)). This technique was compared with the Zephiran decontamination method in use in our hospital. Respiratory specimens (n = 387: 130 endotracheal/bronchial aspirates, 172 bronchoalveolar lavages and 55 sputa) submitted to the University Hospital of Brussels between January 2016 and March 2017 were included. All samples were divided into two aliquots: one was subjected to the Myco-TB method and one to the Zephiran technique prior to culture. The sensitivities for culture for the Zephiran technique on solid media, the Myco-TB method on solid media and Myco-TB combined with the MGIT™ system were respectively 67%, 87% and 89%. The contamination rates were 22% with both the Zephiran and Myco-TB method on solid media and only 4% with the Myco-TB kit combined with the MGIT™ system. For direct microscopy, the sensitivities of the Zephiran method and the Myco-TB method were equal (40%) when the centrifugation time was 20 min. The Myco-TB decontamination method is easy and rapid to perform. It is more sensitive for culture as compared to the Zephiran method and gives lower contamination levels when combined with the MGIT™ technique. When increasing the centrifugation step to 20 min, the sensitivity of direct microscopy is equal to the Zephiran method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.