We studied the anterior tibial compartment pressures during the application of a JOBST sequential intermittent pneumatic compression device on 5 healthy human volunteers (10 legs). Intracompartmental pressures were measured using a slit catheter. The measurements of interstitial pressures were highest at maximal calf inflation, and pressures were increased for approximately 120 seconds during each cycle. Pressure measurements in the inflated pressure sleeve varied less than 10% with the measured anterior tibial compartment pressures during intermittent pneumatic compression therapy. This intermittent pneumatic compression device may elevate intramuscular pressure significantly above that necessary to render muscle ischemic. However, these periods of pressure elevation are not long enough to produce any significant adverse effects, and the beneficial effects of decreased edema fluid may be safely realized.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.