We assess Scanning Thermal Microscopy (SThM) with a self-heated doped silicon nanoprobe as a method for determining the local phase transition temperature of polymeric materials by means of nanothermomechanical analysis (nano-TA). Reference semi-crystalline samples and amorphous test samples, characterized first using differential scanning calorimetry (DSC), are studied by nano-TA in the temperature range 50-250°C. The repeatability, the reproducibility and the reliability of nano-TA are evaluated by three laboratories by applying the same calibration protocol prior to and after the measurements. The calibration of the probe temperature scale and the variability of the sample thermomechanical response are validated by Monte Carlo uncertainty analysis, resulting in a calculated uncertainty between 3 and 5 K. The SThM probe temperature data represented as a function of DSCmeasured phase-transition temperatures of the semi-crystalline samples rule out the possibility of a quadratic fit and call for a linear calibration in absence of additional information. The maximum deviation obtained between SThM and DSC temperatures with such linear calibration reaches ± 30 K for melting temperatures and 50 K for glass transition temperatures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.