Between February 1975 and August 1976, 195 total hip replacements using Stanmore components were performed; of these, 146 were in 135 patients who had not had previous hip surgery. At review 52 had died, but none of the others was lost to follow-up. Of the 52, two had had a second operation, one for infection and one for recurrent dislocation. In the remaining 83 patients (92 hips) five revisions were necessary: four for aseptic loosening, and one for stem fracture. The remaining 78 patients had little or no pain and little restriction of activity. After a follow-up period of at least nine years, the survival rate of the prosthesis was 95%. There had been migration of the femoral component in five cases and migration of the acetabular cup in one case, but no wear of the acetabular component could be demonstrated.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.