BackgroundThe goal of this study was to evaluate the accuracy and interchangeability between continuous cardiac output (CO) measured by electrical velocimetry (COEv) and continuous cardiac output obtained using the pulmonary thermodilution method (COPAC) during living donor liver transplantation (LDLT).MethodTwenty-three patients were enrolled in this prospective observational study. CO was recorded by both two methods and compared at nine specific time points. The data were analyzed using correlation coefficients, Bland-Altman analysis for the percentage errors, and the concordance rate for trend analysis using a four-quadrant plot.ResultsIn total, 207 paired datasets were recorded during LDLT. CO data were in the range of 2.8–12.7 L/min measured by PAC and 3.4–14.9 L/min derived from the EV machine. The correction coefficient between COPAC and COEv was 0.415 with p < 0.01. The 95% limitation agreement was − 5.9 to 3.4 L/min and the percentage error was 60%. The concordance rate was 56.5%.ConclusionsThe Aesculon™ monitor is not yet interchangeable with continuous thermodilution CO monitoring during LDLT.Trial registrationThe study was approved by the Institutional Review Board of Chang Gung Medical Foundation in Taiwan (registration number: 201600264B0).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.