This study presents a novel approach based on a four-electrode electrochemical biosensor for the detection of tau protein -one of the possible markers for the prediction of Alzheimer's disease (AD). The biosensor is based on the formation of stable antibody-antigen complexes on gold microband electrodes covered with a layer of a self-assembled monolayer and protein G. Antibodies were immobilized on the gold electrode surface in an optimal orientation by protein G interaction. Electrochemical impedance spectroscopy was used to analyze impedance change, which revealed a linear response with increasing tau concentrations. The assay is fast (<1 h for incubation and measurement) and very sensitive. The limit of quantification for the full-length 2N4R tau protein is 0.03 pM, a value unaltered when the assay was processed in bovine serum albumin or human serum. This technology could be adapted for the detection of other biomarkers to provide a multiple assay to identify AD progression in a point of care setting.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.