A patient with bilateral implanted neurostimulators suffered significant brain tissue damage, and subsequently died, following diathermy treatment to hasten recovery from teeth extraction. Subsequent MRI examinations showed acute deterioration of the tissue near the deep brain stimulator (DBS) lead's electrodes which was attributed to excessive tissue heating induced by the diathermy treatment. Though not published in the open literature, a second incident was reported for a patient with implanted neurostimulators for the treatment of Parkinson's disease. During a diathermy treatment for severe kyphosis, the patient had a sudden change in mental status and neurological deficits. The diathermy was implicated in causing damage to the patient's brain tissue. To investigate if diathermy induced excessive heating was possible with other types of implantable lead systems, or metallic implants in general, we conducted a series of in vitro laboratory tests. We obtained a diathermy unit and also assembled a controllable laboratory exposure system. Specific absorption rate (SAR) measurements were performed using fibre optic thermometry in proximity to the implants to determine the rate of temperature rise using typical diathermy treatment power levels. Comparisons were made of the SAR measurements for a spinal cord stimulator (SCS) lead, a pacemaker lead and three types of bone prosthesis (screws, rods and a plate). Findings indicate that temperature changes of 2.54 and 4.88 degrees C s(-1) with corresponding SAR values of 9129 and 17,563 W kg(-1) near the SCS and pacemaker electrodes are significantly higher than those found in the proximity of the other metallic implants which ranged from 0.04 to 0.69 degrees C s(-1) (129 to 2471 W kg(-1)). Since the DBS leads that were implanted in the reported human incidents have one-half the electrode surface area of the tested SCS lead, these results imply that tissue heating at rates at least equal to or up to twice as much as those reported here for the SCS lead could occur for the DBS leads.
This paper describes a novel simulator to perform electromagnetic compatibility (EMC) tests for active implantable medical devices (AIMDs) with electromagnetic fields emitted by security systems. The security system simulator was developed in response to over 100 incident reports over 17 years related to the interference of AIMD's with security systems and the lack of a standardized test method. The simulator was evaluated regarding field homogeneity, signal distortion, and maximum magnetic field strength levels. Small three-axis probes and a three-axis scanning system were designed to determine the spatial and temporal characteristics of the fields emitted by 12 different types of walk through metal detectors (WTMDs). Tests were performed on four implanted pacemakers with a saline phantom and correlated to a newly developed test method performed "in air" (without the phantom). Comparison of the simulator thresholds with tests performed in real WTMDs showed that the simulator is able to mimic the pacemaker interference. The interference thresholds found in the simulator indicate that pulsed magnetic fields are more likely to cause interference in pacemakers than sinusoidal fields. The security system simulator will help biomedical engineers, manufacturers of medical devices, and manufacturers of security systems to identify incompatible combinations of WTMDs and AIMDs early in the development stage.
BackgroundThe objective of this study is to investigate electromagnetic compatibility (EMC) of implantable neurostimulators with the emissions from radio frequency identification (RFID) emitters.MethodsSix active implantable neurostimulators with lead systems were tested for susceptibility to electromagnetic fields generated by 22 RFID emitters. These medical devices have been approved for marketing in the U.S. for a number of intended uses that include: epilepsy, depression, incontinence, Parkinsonian tremor and pain relief. Each RFID emitter had one of the following carrier frequencies: 125 kHz, 134 kHz, 13.56 MHz, 433 MHz, 915 MHz and 2.45 GHzResultsThe test results showed the output of one of the implantable neurostimulators was inhibited by 134 kHz RFID emitter at separation distances of 10 cm or less. The output of the same implantable neurostimulator was also inhibited by another 134 kHz RFID emitter at separation distances of 10 cm or less and also showed inconsistent pulsing rate at a separation distance of 15 cm. Both effects occurred during and lasted through out the duration of the exposure.ConclusionsThe clinical significance of the effects was assessed by a clinician at the U.S. Food and Drug Administration. The effects were determined to be clinically significant only if they occurred for extended period of time. There were no observed effects from the other 5 implantable neurostimulators or during exposures from other RFID emitters.
This paper presents results of electromagnetic compatibility (EMC) testing of three implantable neurostimulators exposed to the magnetic fields emitted from several walk-through and hand-held metal detectors. The motivation behind this testing comes from numerous adverse event reports involving active implantable medical devices (AIMDs) and security systems that have been received by the Food and Drug Administration (FDA). EMC testing was performed using three neurostimulators exposed to the emissions from 12 walk-through metal detectors (WTMDs) and 32 hand-held metal detectors (HHMDs). Emission measurements were performed on all HHMDs and WTMDs and summary data is presented. Results from the EMC testing indicate possible electromagnetic interference (EMI) between one of the neurostimulators and one WTMD and indicate that EMI between the three neurostimulators and HHMDs is unlikely. The results suggest that worst case situations for EMC testing are hard to predict and testing all major medical device modes and setting parameters are necessary to understand and characterize the EMC of AIMDs.
The aim of this study was to investigate the mechanisms by which the radiated radiofrequency (RF) GSM (global system for mobile communication) signal may affect pacemaker (PM) function. We measured the signal at the output of the sensing amplifier of PMs with various configurations of low-pass filters. We used three versions of the same PM model: one with a block capacitor which short circuits high-frequency signals; one with a ceramic feedthrough capacitor, a hermetically sealed mechanism connecting the internal electronics to the external connection block, and one with both. The PMs had been modified to have an electrical shielded connection to the output of the sensing amplifier. For each PM, the output of the sensing amplifier was monitored under exposure to modulated and non-modulated RF signals, and to GSM signals (900 and 1800 MHz). Non-modulated RF signals did not alter the response of the PM sensing amplifier. Modulated RF signals showed that the block capacitor did not succeed in short circuiting the RF signal, which is somehow demodulated by the PM internal non-linear circuit elements. Such a demodulation phenomenon poses a critical problem because digital cellular phones use extremely low-frequency modulation (as low as 2 Hz), which can be mistaken for normal heartbeat.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with đź’™ for researchers
Part of the Research Solutions Family.