Radiofrequency (RF) ablation of pulmonary veins (PVs) is a new treatment for atrial fibrillation. Low energy ablation is usually used for this procedure. The effect of superfusate flow on lesion formation in this setting has not been studied previously. We examined lesion dimensions and intramural temperatures with varying powers and duration of RF application in this high flow environment. Ablation of fresh bovine hearts was performed with a 4-mm tip RF catheter in temperature control mode, target temperature 50 degrees C. At power levels of 20 W, 30 W, 40 W, and 50 W, effects of PV flow (no flow or 1 L/min) and 60- and 120-second durations were tested. Tissue temperatures were recorded at depths of 1, 4, 7, and 10 mm. Without flow, no lesions were created. The lowest power setting for lesion creation was 30 W at 60 seconds and 20 W at 120 seconds. Increasing power from 30 W to 50 W for 60 seconds increased lesion depth 0.7 mm (SE 0.3), P = 0.03 and 2.5 mm (SE 0.6), P = 0.003, at 120 seconds. Increasing RF application duration from 60 to 120 seconds increased depth for 30 W by 0.9 mm (SE 0.5), P = NS, 40 W 1.7 mm (SE 0.4), P = 0.002, and 50 W 2.6 mm (SE 0.5), P < 0.001. Power of 50 W for 60 seconds and >30 W for 120 seconds created lesions deeper than the wall thickness of a PV. Flow is necessary for creation of lesions with low power, low tip temperature RF ablation. When a resistant site to ablation is encountered, increasing duration of ablation is best for increasing lesion depth. Higher power has the potential to create lesions deeper than the PV wall and may increase the risk of complications.
Irrigated Nd:YAG laser can be used to make deep narrow myocardial lesions without crater formation. Laser ablation may be more suitable than radiofrequency ablation for intraoperative or catheter-based cure of atrial fibrillation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.