We identified characteristics of devices that caused needle-stick injuries in a university hospital over a 10-month period. Hospital employees who reported needle sticks were interviewed about the types of devices causing injury and the circumstances of the injuries. Of 326 injuries studied, disposable syringes accounted for 35 percent, intravenous tubing and needle assemblies for 26 percent, prefilled cartridge syringes for 12 percent, winged steel-needle intravenous sets for 7 percent, phlebotomy needles for 5 percent, intravenous catheter stylets for 2 percent, and other devices for 13 percent. When the data were corrected for the number of each type of device purchased, disposable syringes had the lowest rate of needle sticks (6.9 per 100,000 syringes purchased). Devices that required disassembly had rates of injury of up to 5.3 times the rate for disposable syringes. One third of the injuries were related to recapping. Competing hazards were often cited as reasons for recapping. They included the risk of disassembling a device with an uncapped, contaminated needle and the difficulty of safely carrying several uncapped items to a disposal box in a single trip. New designs could provide safer methods for covering contaminated needles. Devices should be designed so that the worker's hands remain behind the needle as it is covered, the needle should be covered before disassembly of the device, and the needle should remain covered after disposal. Such improvements could reduce the incentives for recapping needles and lower the risk of needle-stick injuries among health care workers.
These results suggest that shielded syringes and related technology may offer potential for reducing the risk of needlestick injuries among hospital and other healthcare employees. This study should be viewed as a preliminary effort to assess the potential of shielded-syringe technology for reducing the risk of needlestick injuries among healthcare workers. The results are encouraging, but more clinical experience with this new device and longer follow up are essential to provide the data necessary for a definitive assessment of efficacy.
A major factor in the introduction of new products designed to decrease the risk of needlesticks to healthcare workers (HCWs) is whether the increased expense of a safer device is offset by the savings of preventing needlesticks. The itemized costs of needlestick injuries associated with six major needled devices were estimated and compared to the cost of the devices causing the injuries, based on 1988 dollars. Included was the cost of treatment, prophylaxis and employee health department personnel time. The average cost of needlestick injury was $405, with a narrow range of $390 to $456 for different devices. As a percent of the cost of the devices, needlesticks cost as little as 10% of the cost of the device, for the intravenous (IV) catheter, to as much as 457%, for needles used to connect IV lines. On the average, needlesticks cost 36% of the devices' cost. These data may be used to weigh the potential economic benefits of safer needle technology or other strategies intended to reduce the incidence of needlesticks.
A major factor in the introduction of new products designed to decrease the risk of needlesticks to healthcare workers (HCWs) is whether the increased expense of a safer device is offset by the savings of preventing needlesticks. The itemized costs of needlestick injuries associated with six major needled devices were estimated and compared to the cost of the devices causing the injuries, based on 1988 dollars. Included was the cost of treatment, prophylaxis and employee health department personnel time. The average cost of needlestick injury was $405, with a narrow range of $390 to $456 for different devices. As a percent of the cost of the devices, needlesticks cost as little as 10% of the cost of the device, for the intravenous (IV) catheter, to as much as 457%, for needles used to connect IV lines. On the average, needlesticks cost 36% of the devices' cost. These data may be used to weigh the potential economic benefits of safer needle technology or other strategies intended to reduce the incidence of needlesticks.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.