Purpose The purpose of this study was to investigate the spectral response of the EBT3 radiochromic films to different beam qualities for radiation therapy dosimetry. Dose, dose rate, and interbatch dependencies on the spectral response of the films were investigated as well. Methods Pieces of EBT3 films placed between layers of solid water phantoms were irradiated with 6 and 15 MV photon beams, 6 and 10 MV‐flattening filter free (FFF) photon beams, and 6 and 20 MeV electron beams at dose levels between 0.4 and 50 Gy. Net absorbance was measured as a function of wavelength from the spectra acquired in the wavelength range of 400–800 nm using a fiber‐coupled spectrometer and broadband light source. Results No significant change was observed in the absorption spectra of the EBT3 film from the same batch irradiated with the same amount of dose using different beam qualities. Also, no spectral change with dose rate was observed. The measured net absorbance per Gy was independent of beam quality in the 1–50 Gy dose range. Slight differences in the spectral shape and absorption band positions were observed in film samples from different batches. The net absorbance spectra showed two absorption bands centered around 634–636 nm (primary) and 583–585 nm (secondary). However, depending on the film batch, for doses above a certain level the primary absorption band appears to “split” into two bands centered around ~624–628 and ~641–645 nm. Conclusions The spectral shape of the EBT3 radiochromic films irradiated with photons (including FFF) and electron beams is beam quality and dose rate independent; however, it varies with dose level, batch, and spectroscopy system used.
Purpose The purpose of this study was to investigate the beam quality dependence of the EBT‐XD radiochromic films for megavoltage photon and electron beam irradiations by studying their spectral response. Dose, dose rate, and interbatch dependencies were investigated as well. Methods EBT‐XD and EBT3 radiochromic films were cut into 5 cm2 × 5 cm2 pieces, placed between solid water phantoms, and irradiated with 6 and 15 MV photon beams, 6 and 10 MV flattening filter free photon beams, and 6 and 20 MeV electron beams at dose levels between 0.5 and 50 Gy. In order to measure the spectral response, the films were illuminated by a deuterium and tungsten‐halogen lamp and net absorption was measured in 400–800 nm spectral range using a fiber‐coupled optical spectrometer. Film samples were then analyzed using a flatbed scanner in the red, green, and blue channel to obtain the optical density of the films. Results The net absorption spectra of the EBT‐XD and EBT3 films showed two absorption bands centered at 635 and 585 nm, characteristic of the EBT model. However, for a given dose, the EBT‐XD films showed lower net absorbance compared to the EBT3 films. At a given dose level, the net absorption spectra and dose–response curves for EBT‐XD films showed only slight variations across various beam qualities. When a calibration curve obtained from a given beam quality is used to measure dose from films irradiated with other beam qualities, the maximum uncertainty in the calculated dose, in ranges 5–40 and 8–50 Gy for red and green channels, respectively, were ~3.1% and ~2.4%. Conclusions The response of the EBT‐XD films is dose rate independent but batch dependent. No significant beam quality dependence was observed in EBT‐XD films in the dose range up to 50 Gy.
Radiochromic films have been widely studied for clinical dosimetry in conventional external beam radiation therapy. With an increase in practice of proton therapy, such films are being conveniently used; however, their spectroscopic characterization for this modality is lacking. This work investigated the response of the EBT3 radiochromic films irradiated in a Mevion S250™ clinical proton beam. Dose, dose rate, inter-batch, sensitivity, and linear energy transfer (LET) dependencies of the films were studied. Pieces of the radiochromic films from different batches were irradiated using a spread-out Bragg peak (SOBP) proton beam at dose levels between 0.1–15 Gy. Absorption spectra were measured in the wavelength range of 400–800 nm with 2.5 nm resolution. For comparison, the optical density of the films was measured using a flatbed scanner. The net absorbance spectra showed two characteristic absorption bands centered at 636 nm and 585 nm. However, a saturation effect, manifested as broadening/splitting appearance, was observed in the 636 nm band for doses beyond a certain batch-dependent level ~4–10 Gy, in the three different film batches studied. The differences in the spectral shape led to dose-response curves with variable sensitivity. In general a high spectral sensitivity was observed in 0.1–6 Gy range for the three film batches. For a given dose, no significant change in the spectra was observed with change in the dose rate. No significant dependency on the LET was observed for the EBT3 films irradiated with proton beams with dose-averaged LETs ranging from 1.14–6.50 keV µm−1 studied in this work. However, at a given dose, ~5% lower spectral response was observed in the films irradiated with protons compared to their counterparts irradiated with photon beams.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.