We have realized the direct ionization of samples from wells of microtiter plates under atmospheric conditions for mass spectrometry analysis without any liquid delivery system or any additional interface. The microtiter plate is a commercially available 384-well plate without any modification, working as a container and an emitter for electrostatic spray ionization of analytes. The approach provides high throughput for the large batches of reactions and both the qualitative and quantitative analysis of a single compound or mixture. The limits of detection in small drug molecules, peptides, and proteins are similar in comparison with standard direct infusion electrospray ionization. The analysis time per well is only seconds. These analytical merits benefit many microtiter plate-based studies, such as combinatorial chemistry and high throughput screening in enzyme assay or drug metabolism. Herein, we illustrate the application in enzyme assay using tyrosine oxidation catalyzed by tyrosinase in the presence or absence of inhibitors. The potential application in drug development is also demonstrated with cytochrome P450-catalyzed metabolic reactions of two drugs in microtiter plates followed with direct ESTASI-MS/MS-based characterization of the metabolism products.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.