In this work they were developed a microcontroller-based system devoted to the newborn incubator, in order to check the conditions of the environment provided to the premature newborns, a control system of humidity and software that carries out the reading of the sensors. Two microprosecuted, similar newborn incubator were used, for the realization of the measurements, being that one of the newborn incubator used distilled water and other one was valued without water. The measures were collected to each 2 minutes during 96 h. The relative humidity inside of the newborn incubator, without water had average (46.53 ± 3.84) %. The mean value of the internal temperature was of (36.58 ± 1.04) ºC. With water, without control of humidity, (70.67 ± 3.71) %. The mean value of the internal temperature was of (34.33 ± 1.84) ºC. With water and with control of humidity, was inside the established belt of comfort for the standard. One checked that the newborn incubator used in this inquiry does not provide to the premature newborns a totally appropriate environment taking into account the requisites of the standard NBR IEC 601-2-19.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.