Summary: Cholesterol was assayed on the Greiner GSA II and the Technicon SMA 12/60 and SMA II autoanalyzers, vnthAllairfs entirely enzymic method (Allain et al. (1974), Clin. Chem. 20,470-475). The coefficients of variation in the results over a period of one month were less than 6% with the GSA II and the SMA II, but up to 9% on the SMA 12/60. Although the accuracy of the enzymic method depends on the specificity of the cholesterol oxidase and cholesterol esterase, the results were in good agreement with those obtained byAbelPs method (Abell et al. (1952) Allain et al. ((1974), Clin. Chem. 20, 470-475) bestimmt. Die Variationskoeffizienten der ber einen Zeitraum von einem Monat erhaltenen Ergebnisse waren < 6% mit dem GSA II und dem SMA II, aber bis zu 9% mit dem SMA 12/60. Obwohl die Richtigkeit der enzymatischen Methode von der Spezifit t der Cholesterinoxidase und -esterase abh ngt, stimmen die Ergebnisse gut mit denen berein, die mit der Methode von Abell et al. ((1952), J. Biol. Chem. 795, 357-366) an frischem oder lyophilisiertem Plasma von Menschen erhalten wurden. Die Cholestefinkonzentration war bei 4 °C oder in gefrorenem oder lyophilisiertem Plasma stabil. Keine der gepr ften Pharmaka st rten den enzymatischen Test.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.