The possibility of achieving superresolution on a microstep phase image in a laser scanning differential heterodyne microscope is studied both heoretically and experimentally. The superresolution is estimated as the width ratio for the amplitude and phase components of the microscope response measured at half the height of the corresponding parts of the response. It is shown theoretically that superresolution greatly exceeding unity can be achieved for an object in the form of a phase microstep introducing a phase shift equal to π. Superresolution of ∼2 is experimentally obtained for certified test micro-objects. A possibility of tuning a test sample into the superresolution regime by shifting a point photodetector in the microscope’s Fourier plane is demonstrated.
The solution of inverse problem of heterodyne differential microscope based on representation of linear part of microscope response by moments of object function is realized. The parameters of step-like object was determined with superresolution.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.