For the first time, true arterial phase images of the entire carotid artery with high spatial resolution could be acquired using a 16-slice-spiral-CT scanner. This method offers the potential to replace catheter angiography in the evaluation of carotid artery stenosis.
Purpose: Using the BLADE (PROPELLER) technique for T2-weighted MR imaging of the cervical spine has proven to be a reliable tool for reducing artifacts typically for this region. The aim of this study was to evaluate whether the application of BLADE sequences has an impact on the detection of small or low contrast spinal cord and epidural lesions.
Materials and Methods: A standard TSE and a BLADE sequence were compared in 33 patients with 46 spinal cord and epidural lesions for T2-weighted sagittal imaging of the cervical spine. Image sharpness, visualization of the dura, reliability of spinal cord depiction as well as lesion contrast were evaluated by two independent readers. Additionally two experienced neuroradiologists selected in consensus the sequence they would prefer for diagnostic purposes. Statistical evaluations were performed using the sign and the ?2 test.
Results: BLADE was significantly superior to TSE regarding image sharpness, visualization of the dura and reliability of spinal cord depiction. Regarding lesion contrast there was a positive trend towards the BLADE sequence. In 17 of 46 lesions, BLADE was judged superior to TSE, while TSE was favored in 10 lesions. In consensus reading both neuroradiologists preferred BLADE for overall image quality in 27 of 33 patients and for lesion contrast in 10 and TSE in 14 of the 33 patients, but 3 TSE sequences were rated as non-diagnostic regarding this criterion.
Conclusion: For the detection of even small and low-contrast spinal cord lesions, BLADE is at least equivalent to TSE, yielding better overall image quality and fewer non-diagnostic images.
Citation Format:
This hybrid technique in combination with phased-array coils allows a high resolution MRA of the peripheral arteries with very good image quality. If future studies confirm reduced venous overlap, this method may be an alternative also for users of the floating table MRA with the body resonator.
VIBE with iPAT allows MR imaging of the liver with high spatial and temporal resolution providing dynamic contrast-enhanced information about the whole liver. This may lead to improved detection of liver lesions, especially hepatocellular carcinoma.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.