A single crystal and 5 polycrystalline samples of LiF:Mg, Ti and their pellets were prepared and investigated so as to apply thermoluminescence high gamma dose dosimetry. Three zones of single crystal with dopant concentrations of 200 ppm of Mg and 20 ppm of Ti were also used to prepare the single crystal samples. For polycrystalline samples, dopant concentrations of 0.062 mol% Mg and Ti concentrations in the range of 0.016 and 0.046 mol% were used. All the samples were exposed to gamma doses of 1 kGy to 700 kGy and their response changes were determined by a gamma dose test of about 60 mGy. According to the results obtained, the use of response reduction by curve-fitting up to about 300 kGy can be performed reliably for high dose gamma dosimetry.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.